STAR aspirate: CAPACITIVE and PRESSURE search first, ZTOUCH draws from the floor
A liquid height given by hand is only as right as the tracker or the plate's seating. The firmware's own LLD in C0 AS is left off: the channels search first, as probe_liquid_heights does, from the command's lp reached at approach_speed, and the command then starts where the tips rest, at the surface found. The tracker takes the measured volume, with a warning when it is 20 % off the model. ZTOUCH touches the floor first and draws from it. Every search runs 1 mm below the modelled cavity bottom, and a surface found below it moves the floor sent. The command draws its blow-out air before its descent, which with the tip already on the liquid would draw liquid. Under an LLD mode that air is drawn beforehand, in air, by Px DC, and C0 AS carries none. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>